EUTECH INSTRUMENTS 2700 Manual de instrucciones Pagina 35

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 60
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 34
15. Dissolved Oxygen Setup
Use Setup Dissolved Oxygen mode to customize the parameter. Press
from the measurement screen and then press ENTER when DO% or DO (mg/L)
is selected for percent saturation or concentration modes respectively. Note that
setup parameters differ slightly depending on which is selected.
Sample ID
This is a user selectable number from 1 to 99999.
Incorporating a sample ID to identify one or more
data points is useful to distinguish data that is
saved into memory or sent to a PC or printer. Use
the up/down arrow keys to adjust the values and
left/right arrow keys to move the cursor.
Offset Cal (% saturation only)
Use this feature to offset the meter’s value to
another DO meter. Using the % saturation mode,
observe the reading a sample solution after it has
stabilized. Similarly, observe the reading of the
same sample using another DO meter as a
reference. The probe of the reference meter should
be immersed in the same sample at the same
depth.
The DO 2700 allows ±10% offset adju
stment. Note that subsequent user
calibrations will reset the offset adjustment back to the factory default value of 0.0
% offset.
Pressure Unit (% saturation mode only)
Select mmHg or kPa pressure units.
31
Vista de pagina 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 59 60

Comentarios a estos manuales

Sin comentarios