Operating Instructions αlpha-CON1000
10
5.2.2.1 Clearing the Advanced Setup security code from the display
After you have entered the security code and returned to the Measurement mode, the security code
“22” still appears on the display whenever you press the ENTER key. To conceal the security code,
you must manually reset the code. To clear the Advanced Setup security code from the display:
1) Press the ENTER key in the Measurement mode.
2) Set to any security code (not 11 or 22) and complete by pressing ENTER.
NOTE: When you enter the Calibration mode with code “11” or Advanced Setup mode with
security code “22”, the unit automatically enters into the HOLD mode until you return back to
Measurement mode. The HOLD annunciator is displayed at the upper left of the display. While on
HOLD, the current output is frozen and set point relays are deactivated.