A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
4 6 0 0 S E R I E S T H E R M O M E T E R
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
4 6 0 0 S E R I E S T H E R M O M E T E R
7
Example #1: You want F1 to be the F/C function,
and F2 to be the HOLD function.
Turn on the unit.
After the unit has initialized, press the MODE key
down 5 times (C or F will be flashing).
Press F1. (F1 is now the degrees C/F function key.
The thermometer has also changed from C to F or
from F to C.)
Now, assign HOLD to the F2 key.
Press MODE once. (HOLD is flashing)
Press F2. (F2 is now the HOLD function. The
thermometer has also been put into the HOLD
mode. Press F2 again to go back to normal mode.)
Example #2: You want to t a k e de lta T
measurements. (The difference between a
val u e you have store d , us ing the STOR E
function, and the meters present reading. In
this example, let us assume that you are taking
the temperature d i f f e r e n c e b e t w e e n two
beakers of water, beaker A and beaker B.)
- O P E R A T I N G I N S T R U C T I O N S -
5. KEY PAD OPERATION:
20
- S P E C I F I C A T I O N S -
TEMPERATURE
(°C) (°F)
-40 (-40)
-30 (-22)
-20 (-4)
-10 (14)
0 (32)
10 (50)
20 (68)
30 (86)
40 (104)
50 (122)
60 (140)
70 (158)
80 (176)
90 (194)
100 (212)
110 (230)
120 (248)
130 (266)
140 (284)
150 (302)
ACCURACY
(°C) (°F)
0.100 (0.180)
0.050 (0.090)
0.025 (0.045)
0.020 (0.036)
0.015 (0.027)
0.015 (0.027)
0.015 (0.027)
0.015 (0.027)
0.015 (0.027)
0.015 (0.027)
0.020 (0.036)
0.025 (0.045)
0.030 (0.054)
0.045 (0.081)
0.065 (0.117)
0.080 (0.144)
0.110 (0.198)
0.150 (0.270)
0.240 (0.432)
0.320 (0.576)
MODEL 4600-1 & 4600S* SPECIFICATIONS:
INSTRUMENT ACCURACY
VS. TEMPERATURE**
(Not including probe error)
**Ambient temperature: 18 to 28°C (64 to 82°F)
*The 4600S is a standard 4600 which is calibrated
at a specific temperature or temperatures, up to
four. The thermometer and probe are calibarted as a
system and must stay together to maintain the system
level accuracy specification shown on page 19.
Model 4600 Manual-3.5x8.5-rev.indd, Spread 10 of 16 - Pages (10, 23) 1/11/2007 2:03 PM