EUTECH INSTRUMENTS ALPHA COND 500 Especificaciones Pagina 16

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 32
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 15
15
5.3 Conductivity Calibration
A 1-point Calibration is required for this transmitter. If
the calibration process is aborted, your transmitter
will revert to the previous calibration data.
1. Organize your calibration standard solution
in two beakers – one for rinsing and the
other for calibration. Prepare separate de-
ionized water for electrode rinsing.
2. Rinse the electrode in de-ionized water and
then rinse with the calibration standard.
3. Dip the electrode to the calibration standard
intended for calibration and swirl gently to
create a homogenous sample. Wait for the
reading to stabilize.
4. Press the CAL key to enter calibration
mode. The upper display will show the
measured reading based on the last
calibration setting whereas the lower display
will show the measured default reading of
the standard.
5. Use the or keys to adjust the upper
display to the calibration standard value.
Note: You can offset the conductivity reading within
± 40% from the default reading of the standard. The
minimum allowable calibration point is 10% of the full
scale reading of the range selected.
6. Press the ENTER key to confirm the calibration value and to return to the
measurement mode.
Note: To exit this program without confirming the calibration, press the and
keys together (Escape).
Note: When there is a calibration error, the buffer icon and ERR annunciator will
appear together with a blinking electrode icon. Press both and keys together
TWICE to return to the measurement mode.
HOLD
HOLD
MEAS
°C
1578
25.5
ATC
μ
S
CAL
1578
1578
μ
S
CAL
1413
1578
μ
S
Vista de pagina 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 31 32

Comentarios a estos manuales

Sin comentarios