Instruction Manual alpha-pH 1000
28
1) Follow directions in 7.6.1 to enter Configuration mode.
2) Press the ENTER key until the upper display shows “no”, “FCt” or “CAL”,
and the lower display shows “deF” (default).
3) Press the or key to select “no” for using the old values, “FCt” to reset
all settings to factory defaults or “CAL” to reset calibration values only to the
factory default values.
WARNING: Changing from “no” to “FCt” or “CAL” and pressing the ENTER key
resets all settings or the calibration values respectively to factory defaults.
4) Press the ENTER key to confirm your selection and to return to Advanced
Set-up mode, or return to Measurement mode by pressing the and
keys simultaneously (escape).
7.7 Calibration (CAL) sub-function
The calibration procedure in Advanced Set-up mode is identical to the procedure in the Calibration mode. The
only difference is that the controller will revert back to Set-up mode (instead of Measurement mode) after
calibration is completed.
7.7.1 Entering Calibration mode from Advanced Set-up mode
1) Enter Advanced Set-up mode. Push the ENTER key and scroll to Advanced Set-up security code “22”.
Push the ENTER key again.
2) Press the or key to scroll until the upper display shows “CAL”.
3) See section 6 for complete calibration procedures.
no
dEF
SETUP
HOLD
FCt
dEF
SETUP
HOLD
CAL
dEF
SETUP
HOLD