  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 25
Calibration error.
Buffer value does not
match displayed value or
electrode is disconnected
or failing.
Use fresh buffer solutions.
Check electrode connection.
Clean & recondition electrode.
Replace electrode.
Broken or worn electrode Replace electrode Unstable pH
reading
External ‘noises’ or
induction (electrical ‘noise’
caused by a nearby motor)
Remove or switch off interfering
device
Slow
response
Dirty electrode. Clean & rehydrate the electrode
CAL
during pH
CAL mode
Calibration was attempted
using only 1-point that was
not pH 7 (USA) or pH 6.86
(NIST).
Recalibrate using 2 or more pH
points or use pH 7 (USA)
or use pH 6.86 (NIST).