EUTECH INSTRUMENTS DO 6 PLUS Manual de instrucciones Pagina 15

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 48
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 14
Instruction Manual DO 6+
- 9 -
The DO 6+ will accept one calibration point in concentration mode. The
minimum value is 2 mg/L (ppm), and the calibration window is +/- 40% of the
factory default value.
Temperature and % saturation calibration should take place before attempting
to perform mg/L (ppm) concentration calibration.
Calibration of the concentration mode will only replace the previous
concentration calibration and does not affect the % saturation calibration.
To offset your % saturation reading to match another instrument, see Section
5.8 — % Saturation Offset Adjustment.
New calibration values will automatically override the existing data. Perform
daily calibration for best results. To completely recalibrate the meter and when
installing a replacement electrode, it is best to clear all calibration data. See
Section 5.7Reset to Factory Default.
Vista de pagina 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 47 48

Comentarios a estos manuales

Sin comentarios