
Instruction Manual Salt 6
19
6 ADVANCED SETUP FUNCTIONS
6.1 Advanced Setup Overview
The advanced setup mode lets you customize your meter’s preferences and
defaults. To enter advanced setup mode:
1) Make sure that the meter is switched-off.
2) Press ON and MODE key simultaneously, holding both keys for 2
seconds. First release ON key first before releasing the MODE key.
3) [StUP] indicator will appear momentarily and [tCo%] will appear
next.
Enter Setup Page.
%
Adjust Temperature Coefficient value from 0.0
to 3.0 %/°C.
Default value is 2.1 %/°C.
C°
Select Normalization Temperature. Choice of
either 20 °C or 25 °C.
Default value is 25 °C.
To choose measurement parameter, TDS (ppt)
or percentage (%).
User re-set to factory defaults. Choice of “Yes”
or “No”.
Default value is “no”.
Comentarios a estos manuales