page | 77
[ Continued from page 76 ]
Models PCD 650 PC 650 PD 650 CD 650 PD 300 PC 300 PC 10
Multi-Parameter
Handheld Meters
Specications
Salinity
Range
.. to 0.770 ppm
0.770 ppm to 143.3 ppm
143.3 ppm to 2.138 ppt
2.138 ppt to 23.64 ppt
23.64 ppt to 80.00 ppt
–
.. to 0.770 ppm
0.770 ppm to 143.3 ppm
143.3 ppm to 2.138 ppt
2.138 ppt to 23.64 ppt
23.64 ppt to 80.00 ppt
–
Resolution 0.01 ppm / 0.1 ppm / 0.001 ppt / 0.01 ppt –
0.01 ppm / 0.1 ppm /
0.001 ppt / 0.01 ppt
–
Accuracy ±1 % full scale + 1 LSD – ±1 % full scale + 1 LSD –
Resistivity
Range
2.000 Ω to 25.00 Ω
25.00 Ω to 250.0 Ω
250.0 Ω to 3.333 kΩ
3.333 kΩ to 500.0 kΩ
500.0 kΩ to 20.00 MΩ
–
2.000 Ω to 25.00 Ω
25.00 Ω to 250.0 Ω
250.0 Ω to 3.333 kΩ
3.333 kΩ to 500.0 kΩ
500.0 kΩ to 20.00 MΩ
–
Resolution 0.01 Ω / 0.1 Ω / 0.001 Ω / 0.1 Ω / 0.01 MΩ –
0.01 Ω / 0.1 Ω / 0.001Ω /
0.1 Ω / 0.01 MΩ
–
Accuracy 1 % full scale – 1 % full scale –
Dissolved
Oxygen
Range
0.00 to 90.00 mg/L
or ppm
– 0.00 to 90.00 mg/L or ppm
0.00 to 20.00 mg/L
or ppm
–
Resolution 0.01 mg/L ; 0.01 ppm – 0.01 mg/L ; 0.01 ppm –
Accuracy ±0.2 mg/L – ±0.2 mg/L ±1.5 % full scale –
Cal. Points 1 – 1 –
Probe
Galvanic with integral
temperature sensor
–
Galvanic with integral
temperature sensor
–
% Saturation
of Oxygen
Range 0 to 600.0 % – 0 to 600.0 % 0.0 to 200.0 % –
Resolution 0.1 % – 0.1 % 0.1 % –
Accuracy ±2.0 % – ±2.0 % ±1.5 % full scale –
Salinity
Correction
Range 0.0 to 50.0 ppt – 0.0 to 50.0 ppt 0.0 to 50.0 ppt –
Resolution 0.1 ppt – 0.1 ppt –
Method Automatic / manual –
Automatic
correction after
manual input
Automatic / manual
Automatic
correction after
manual input
–
Barometric
Pressure
Correction
Range
450 to 825 mmHg
or 59.9 to 109.9 kPa
–
450 to 825 mmHg
or 59.9 to 109.9 kPa
500 to 1499 mmHg
or 66.6 to 199.9 kPa
–
Resolution 0.1 mmHg or 0.1 kPa – 0.1 mmHg or 0.1 kPa –
Method
Automatic
correction with
built-in sensor
– Automatic correction with built-in sensor
Automatic
correction after
manual input
–
Meter
Features
Temperature
Compensation
ATC / MTC (0 to 100 ºC)
GLP Yes –
Cal-Due Alarm Yes –
IP67 Yes
Datalogging Yes –
Memory 500 data sets 50 data sets –
Operating
Temperature
0 to 50 ºC
Ready Function Yes
Hold Function Yes
LCD Display
Dot-matrix LCD with backlight
(5.4 x 7.1 cm)
Dual-display LCD
with backlight
(5.8 x 3.3 cm)
Dual-display LCD
(5.8 x 3.3 cm)
Auto-Off 2 to 30 mins after last key pressed 20 mins after last key pressed
Input
DC phono sockets,
BNC, 8-pin
connector,
6-pin connector
DC phono sockets,
BNC,
8-pin connector
DC phono sockets,
BNC, 8-pin
connector
6-pin connector
DC phono sockets,
6-pin connector,
8-pin connector
BNC, 6-pin connector 6-pin connector
Output IrDA, RS232C (via LED) * –
Power 4 x 1.5 V ‘AA’ alkaline batteries or 9 V DC adapter, 500 mA 4 x 1.5 V ‘AAA’ alkaline batteries
Battery Life > 200 hrs > 400 hrs > 200 hrs > 100 hrs > 50 hrs
Dimensions
(LxWxH)
; Weight
Meter 18.3 x 9.5 x 5.7 cm ; 460 g 19 x 10 x 6 cm ; 700 g 19 x 10 x 6 cm ; 320 g
Boxed 40 x 33 x 10 cm ; 3200 g 40 x 33 x 10 cm ; 2180 g
* RS232C (LED) interface adapter available as separate accessory (see page 78 for order information)