Instruction Manual Alpha DO 500
30
5.8 Viewing Calibration Point
Calibration setting lets you view the high-point at which the calibration has been
performed in the Transmitter.
ENT
SETUP
HOLD
1
ENT
SETUP
HOLD
2
%
11
From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode
(COF). Press  or  key to select viewing calibration settings screen
(CAL).
Press ENT key to view calibration settings (CAL).
2
The buffer annunciator
appears in the LCD. The upper display shows
the value of the high-point at which the calibration has been carried out.
The corresponding unit of measurement of calibration is displayed.
NOTE:
If the Transmitter has not been calibrated yet for the currently
selected unit of measurement, the LCD shows an empty
reading (“---“ ) in the upper display.
Press ENT key to exit calibration settings screen. The Transmitter
reverts to SET CAL screen.
Press  or  key to access other setup screens or press  and  key
simultaneously (escape) to return to measurement mode.