Page | 10
A General purpose, aqueous pH measurements. Quality control,
general laboratory and aquarium water.
B Tris buffers, clinical, and biological media containing proteins,
creams, fats and cosmetics.
C Research measurements, fruit juices, beer, milk and yogurt.
D High viscosity solutions such as emulsions, suspensions, paints
and varnishes where frequent cleaning of the reference is
necessary.
E Non aqueous solutions. Must be used with special reference
electrolyte ECRE015.
F Low temperature measurements.
G Low ionic strength solutions, such as high purity water.
H High temperature measurements.
I Surface measurements such as paper, skin, textile, leather and
agar plates.
J Solid or semi-solid samples such as cheese, meats, fruits, bread,
or other similar sample.
K Rugged use; ruggedized electrode tip. Aqueous media in general.
L Portable pH meter; field and school use.
M Micro sample, such as semi-micro cuvettes and NMR tube.
N Test tube measurements.
O Reference contamination problems which require a double
junction reference, such as media containing sulfides, bromides,
iodides, and cyanides.
P Measurements in long, narrow vessels.
Q Soil pH measurements.
R Swimming pool, pH control under continuous flow conditions.
S Isoelectric focusing gels and other surface measurements that
require a small diameter flat tip.
T Photographic chemical; high pH sample (12 to 14 pH).
U ORP measurements for general purpose.
V Swimming pool ORP measurement.
W ORP measurement for cyanide use.
X ORP measurement requiring a double junction reference.
Y Hydrofluoric acid and abrasive solution measurements.
Z Potentiometric measurements of chlorides and other
halides.
Membrane Glass Types
X-1 is especially suited for low temperature, nonaqueous solutions and pH measurements under pH 12.
+
X-2 is best suited for most pH measurements where minimal or no Na is present. It is a low-resistant glass
with a very fast and stable response and is designed for pH ranges of 0 to 13 and temperatures of 0 °C to
135°C.
X-3 is especially formulated for continuous, long-term use at high temperatures, particularly in strong alkaline
+
solutions above pH 11. It experiences negligible Na ion error above 13 pH. The impedance of the glass is
much higher than other glasses, and a slower response will be experienced at room temperature and below.
Response time increases as the temperature is elevated.
Electrodes | Electrode Selection Guide
Electrode Selection Guide
4 Simple Steps To Choosing Your Electrode
Step 1: Select application Step 3: Select electrode material
Step 4: Select from table
Step 2: Select membrane glass type of your electrode
Application Codes & Descriptions
A General purpose, aqueous pH measurements. Quality
control, general laboratory and aquarium water.
B Tris buffers, clinical, and biological media containing
proteins, creams, fats and cosmetics.
Membrane Glass Types
X-1 is especially suited for low temperature, nonaqueous
solutions and pH measurements under pH 12.
X-2 is best suited for most pH measurements where
minimal or no Na+ is present. It is a low-resistant glass
and is designed for pH ranges of 0 to 12 and
0 0
temperatures of 0 C to 135 C
Glass body Combination pH Electrodes
Order Code Glass TYPE Application Code pH Range Temperature
Range(0C)
ECFG73504 X-2 A, F 0 - 13 0 to 100
ECFG73905 X-2 B, C, G 0 - 13 -5 to 80
ECFG73792 X-2 H, O 0 - 13 0 to 110
Order Code Glass TYPE Application Code pH Range Temperature
Range(0C)
ECFG73593 X-2 H 0 - 13 0 to 100
ECFG79701 X-2 E, O 0 - 13 -0 to 100
ECFG74519 X-3 H, T 0 - 14 5 to 110
ECFG63506 X-2 A, D 0 - 13 0 to 100
Order Code Glass Type Application Code pH Range Temperature Range (ºC) Length x Dia.(mm) Internal Reference Element
ECFE74526 X-3 T 0 - 14 -5 to 80 115 x 12 Ag/AgCl Sealed
ECFE72511 X-1 I 0 - 12 0 to 100 115 x 12 Ag/AgCl Sealed
ECFE53901 X-2 B, N, P 0 - 13 -5 to 80 155 x 6 Calomel Sealed
ECFC72521 X-1 A, L, R 1 - 13 0 to 80 110 x 12 Ag/AgCl Sealed
ECFC72521R X-1 A, L,R 1 - 13 0 to 80 110 x 12 Ag/AgCl Refillable
ECFC72522 X-1 A, L, O, R 1 - 13 0 to 80 110 x 12 Ag/AgCl Double Junction Sealed
ECFC72522R X-1 A, L, O, R 1 - 13 0 to 80 110 x 12 AgAgCl Double Junction Refillable
EC620132 X-2 A, B, G, L, O, Q 0 - 14 0 to 60 120 x 12 Ag/AgCl Double Junction Sealed
EC620109 X-2 B, D, G, L, Q 0 - 14 0 to 100 175 x 12 Ag/AgCl Double Junction Refillable
EC620116 X-2 A, B, D, G, L, Q 0 - 14 0 to 100 175 x 12 Ag/AgCl Double Junction Refillable
EC620300 X-2 A, B, L 0 - 14 -5 to 80 106 x 10 Calomel Refillable
Plastic and Epoxy Combination pH Electrodes
Order Code Glass Type Application Code pH Range Temperature Range (ºC) Length x Dia.(mm) Internal Reference Element
ECFG73504 X-2 A, F 0 - 13 0 to 100 110 x 12 Ag/AgCl Refillable
ECFG73905 X-2 B, C, G 0 - 13 -5 to 80 110 x 12 Calomel Refillable
ECFG73792 X-2 H, O 0 - 13 0 to 110 110 x 12 Ag/AgCl Double Junction Sealed
ECFG73593 X-2 H 0 - 13 0 to 100 110 x 12 Ag/AgCl Sealed
ECFG73701 X-2 E, O 0 - 13 0 to 100 110 x 12 Ag/AgCl Double Junction Refillable
ECFG74519 X-3 H, T 0 - 14 5 to 110 110 x 12 Ag/AgCl Sealed
ECFG63506 X-2 A, D 0 - 13 0 to 100 55 x 8 Ag/AgCl Refillable
ECFG43902 X-2 B, C, M 0 - 13 -5 to 80 180 x 4 Calomel Refillable
ECFG43904 X-2 B, C, M 0 - 13 -5 to 80 55 x 4 Calomel Refillable
ECFG73521 X-2 Q 0 - 13 0 to 100 110 x 12 Ag/AgCl Refillable
ECFG63511 X-2 J 0 - 13 0 to 100 25 x 8 Ag/AgCl Sealed
ECFG53912 X-2 B, J 0 - 13 -5 to 80 25 x 6 Calomel Refillable
ECFG72520 X-1 I 0 - 12 0 to 100 110 x 12 Ag/AgCl Refillable
ECFG73711 X-2 D, O 0 - 13 0 to 100 110 x 12 Ag/AgCl Double Junction Refillable
ECFG73511 X-2 D 0 - 13 0 to 100 110 x 12 Ag/AgCl Refillable
ECFG72710LI X-2 E 0 - 13 0 to 100 110 x 12 LiCl Double Junction Refillable
EC620130 X-2 A, B, C, G, L, O, Q, R 0 - 14 0 to 80 160 x 12 Ag/AgCl Double Junction Refillable
EC620131 X-2 A, B, C, D, L, O, Q, R 0 - 14 0 to 80 130 x 12 Ag/AgCl Double Junction Sealed
EC620133 X-2 A, B, I, O, Q 2 - 14 0 to 50 25 x 6 Ag/AgCl Double Junction Sealed
EC620185 X-2 A, B, K, L, O 0 - 14 -5 to 100 102 x 10 Ag/AgCl Double Junction Refillable
EC620181 X-2 A, K 0 - 14 0 to 80 102 x 10 Ag/AgCl Refillable
EC620183 X-2 A, B, K 0 - 14 0 to 80 102 x 10 Ag/AgCl Double Junction Refillable
EC620286 X-2 A, B, G 0 - 14 -5 to 80 106 x 10 Calomel Refillable
EC620293 X-2 B, N, P 0 - 14 -5 to 80 150 x 6 Calomel Refillable
EC620095 X-2 B, N, P 0 - 14 0 to 80 254 x 3 Calomel Refillable
EC620096 X-2 B, N, P 0 - 14 0 to 80 127 x 3 Calomel Refillable
EC620297 X-2 B, N, P 0 - 14 -5 to 100 165 x 5 Ag/AgCl Refillable
Glass Body Combination pH Electrodes
All electrodes are supplied with standard cable length of 1m and with BNC connectors. Refer to our price list for electrodes with longer cable length.
Order Code Sensor Type Application Code Temperature Range (ºC) Length x Dia. (mm) Internal Reference Element
ECFG79601 Platinum U 0 to 100 110 x 12 Ag/AgCl Refillable
ECFE77689 Antimony Y 0 to 80 110 x 12 Ag/AgCl Sealed
ECFG78610 Platinum U, X 0 to 100 110 x 12 Ag/AgCl Double Junction Sealed
ECFC79601 Platinum Pin U, V 0 to 80 110 x 12 Ag/AgCl Sealed
ECFC79601R Platinum Pin U, V 0 to 80 110 x 12 Ag/AgCl Refillable
ECFC79602 Platinum Pin U, V, X 0 to 80 110 x 12 Ag/AgCl Double Junction Sealed
ECFE79602R Platinum Pin U, V, X 0 to 80 110 x 12 Ag/AgCl Double Junction Refillable
Plastic, Epoxy and Glass Combination ORP Electrodes
Page | 11
Electrodes | Electrode Selection Guide
Comentarios a estos manuales