Instruction Manual Temp 4/5/6
10
3. CALIBRATION
3.1 Temperature Calibration
The temperature sensor included (only for Temp 5) with your meter is factory
calibrated. Over time, the temperature calibration may drift and the probe
requires recalibration. The Temp 4/5/6 meters allow you to have a 1-point
calibration as fine adjustment by changing its offset value. This is useful if you
replace the probe and should recalibrate the probe to the meter.
3.1.1 Temperature Calibration using Offset Adjustment
1. Connect your temperature probe to the meter. Dip the probe in a constant
temperature bath or a liquid whose temperature can be checked with a
thermometer known to be accurate. For best accuracy, place the probe and
thermometer in a constant temperature bath.
2. Turn on the meter. Make sure that the meter is in the measurement mode.
Press CAL key to enter temperature calibration mode. The LCD shows
CA” momentarily and the displayed reading flashes.
.
°C
3. Press S and T keys to adjust the displayed value until it matches the
correct desired temperature. The S and T key will scroll to the maximum
allowable value (maximum adjustment is ±5
o
C from factory default).
4. Press ENTER key to confirm calibration. The LCD displays “CO
momentarily, and the meter then reverts to measurement mode.
.
°C