Instruction Manual
CON 11 & CON 110
Handheld Conductivity/TDS/Temperature/RS232C Meter
Technolo
g
y
M
ade
E
a
s
y
.
.
.
68X361303
Rev. 2 - 01/04
Vista de pagina 0
1 2 3 4 5 6 ... 79 80

Indice de contenidos

Pagina 1 - CON 11 & CON 110

Instruction Manual CON 11 & CON 110 Handheld Conductivity/TDS/Temperature/RS232C Meter TechnologyMadeEasy...68X361303 Rev. 2 - 01/04

Pagina 2 - Rev 2 - 01/04

Instruction Manual CON 11/110 63.2 Conductivity Electrode Information Your meter includes a conductivity electrode (Part No: EC-CONSEN91W/ 35608-

Pagina 3 - Table of Contents

Instruction Manual CON 11/110 73.3 Connecting the electrode to the meter The CON 11 / 110 meter use the Conductivity / TDS electrode with a sturd

Pagina 4

Instruction Manual CON 11/110 83.4 Attaching the Electrode Holder to the Meter The CON meter is designed to allow you to do a one hand operation

Pagina 5 - 1 INTRODUCTION

Instruction Manual CON 11/110 93.5 Connecting the AC/DC Adapter Besides using the 4 “AAA” alkaline batteries as the power source, the CON 11/110

Pagina 6 - 1.8.8.8

Instruction Manual CON 11/110 103.6.1 RS232C Configuration The CON 110 meter has a 9 pin female RS232C connector with the following pin out. PIN

Pagina 7 - 2.2 Keypad

Instruction Manual CON 11/110 114 CALIBRATION 4.1 Important Information on Meter Calibration Your meter has five measuring ranges. You have an op

Pagina 8

Instruction Manual CON 11/110 124.2 Preparing the Meter for Calibration Before starting calibration, make sure you are in the correct measurement

Pagina 9 - MADE IN S'PORE

Instruction Manual CON 11/110 134.3 Temperature Calibration Your electrode features a built-in temperature sensor. The temperature sensor is facto

Pagina 10 - 35608-50)

Instruction Manual CON 11/110 144.4 Selection of Automatic or Manual Calibration This meter is capable of performing either automatic conductivity

Pagina 11

Instruction Manual CON 11/110 154.5 Automatic Calibration (For Conductivity Calibration Only) In the Automatic Calibration mode, the meter is cap

Pagina 12 - Figure 9: One hand

Preface This manual serves to explain the use of the CON 11/110 Conductivity/TDS handheld meters. It functions as a handy reference step by step guid

Pagina 13

Instruction Manual CON 11/110 16 IMPORTANT NOTES: 1. Meter allows a tolerance range of ±40% of its calibration standard. If the calibration stan

Pagina 14

Instruction Manual CON 11/110 174.6 Manual Calibration (For Conductivity & TDS Calibration) In Manual Calibration mode, you can use customise

Pagina 15 - 4 CALIBRATION

Instruction Manual CON 11/110 184.7 TDS Calibration 4.7.1 Calibrating for TDS Using Conductivity Standards & adjusting TDS factor It is impor

Pagina 16

Instruction Manual CON 11/110 19 4.7.3 Calibrating for TDS using TDS standards After setting the correct TDS Factor, you can commence calibration

Pagina 17 - 4.3 Temperature Calibration

Instruction Manual CON 11/110 205 MEASUREMENT This meter is capable of taking measurements with automatic (ATC) or manual (MTC) temperature compen

Pagina 18

Instruction Manual CON 11/110 215.2 Manual Temperature Compensation To select the Manual Temperature Compensation (MTC) option, set up the Manual

Pagina 19 - Conductivity Calibration

Instruction Manual CON 11/110 225.2.1 Setting the manual temperature compensation value To use manual temperature compensation (MTC), you need to

Pagina 20

Instruction Manual CON 11/110 235.3 Taking Measurements 1. Rinse the electrode with de-ionised or distilled water before use to remove any impuri

Pagina 21 - Calibration

Instruction Manual CON 11/110 245.4 Using Manual Ranging Function Your meter automatically selects the range in which your readings appear. The m

Pagina 22 - 4.7 TDS Calibration

Instruction Manual CON 11/110 255.5 HOLD Function This feature lets you freeze the display for a delayed observation. HOLD/ENTER can be used any t

Pagina 23

Table of Contents 1 INTRODUCTION...

Pagina 24 - 5 MEASUREMENT

Instruction Manual CON 11/110 266 MEMORY FUNCTION 6.1 Memory Input CON 11/110 meter stores up to 50/100 set of data respectively. Data sets can i

Pagina 25 - Compensation Set Up sequence

Instruction Manual CON 11/110 276.2 Memory Recall This function recalls the previous readings stored in the memory. You can access memory recall f

Pagina 26 - Compensation Value

Instruction Manual CON 11/110 287 PRINT FUNCTION (FOR CON 110 ONLY) The CON 110 meter is equipped with an RS232 communication port and a PRINT (

Pagina 27 - 5.3 Taking Measurements

Instruction Manual CON 11/110 297.2 Sending Data To Computer or Printer When the PRINT ( ) key is pressed, there are two options: 1. dAt / CUr:

Pagina 28 - MEAS Blinking

Instruction Manual CON 11/110 307.2.1 Printing Current Data Manually To manually print any data that is currently being measured: 1. From measu

Pagina 29 - 5.5 HOLD Function

Instruction Manual CON 11/110 31 5 10secsecFrom measurement mode"MEM" annunicator will keep flashing during measurement mode & data

Pagina 30 - 6 MEMORY FUNCTION

Instruction Manual CON 11/110 327.2.3 Print Data from Stored Memory The CON 110 meter can print data that is stored in the memory. You can print

Pagina 31 - 6.2 Memory Recall

Instruction Manual CON 11/110 33To print Selected Stored Data 1. From measurement mode, press MR/▼ key. 2. A “MEM” indicator appeared and the la

Pagina 32

Instruction Manual CON 11/110 348 ADVANCED SETUP FUNCTION The advanced setup mode lets you customised your meter’s preferences and defaults. The

Pagina 33

Instruction Manual CON 11/110 35 MIMRSETUPP 6.0CELSETUPP 8.0AscSETUPP 9.0SETUPP 7.0SETUPP 5.0MIMIMIMRMRMRMIMRSETUPMEM P 1.0SETUPCAL P 2.0SETUPEL

Pagina 34 - Current data manually

8.3 P2.0: VIEWING CALIBRATION DATA... 42 8.4 P3.0

Pagina 35 - From measurement mode

Instruction Manual CON 11/110 368.1 Advanced SETUP Mode Overview 1. Press the SETUP key to enter Set Up menu. 2. Press the MI/▲ or MR/▼ keys to

Pagina 36 - To stop, press PRINT key

Instruction Manual CON 11/110 37CELSETUP P 8.0 P8.0: Selection of cell constant P8.0 Selection of cell constant: K = 1.0, 10, or 0.1 AscSETUP

Pagina 37 - Repeat print procedure or

Instruction Manual CON 11/110 388.2 P1.0: Unit Configuration (COF) 8.2.1 P1.1: Selection of READY and Auto HOLD function Program P1.1 gives you

Pagina 38 - 8 ADVANCED SETUP FUNCTION

Instruction Manual CON 11/110 398.2.2 P1.2: Selection of °C or °F Available in CON 110 only You can select between °C and °F units for temperatu

Pagina 39

Instruction Manual CON 11/110 408.2.3 P1.3: Selection of Automatic or Manual Temperature Compensation This feature lets you select between Automa

Pagina 40

Instruction Manual CON 11/110 418.2.4 P1.4: Selection of TDS Conversion Factor You can set the TDS conversion factor between 0.40 and 1.00; meter

Pagina 41

Instruction Manual CON 11/110 428.3 P2.0: Viewing calibration data This mode lets you recall the previous calibration data on each of the five me

Pagina 42 - OFF or HOLD

Instruction Manual CON 11/110 438.4 P3.0 Viewing Electrode Data Program 3.0 has five “view only” options that let you check the electrode’s param

Pagina 43 - °C or °F

Instruction Manual CON 11/110 448.5 P4.0: Auto Off In Program P4.0 you can activate the Auto Off option to automatically switch off the instrume

Pagina 44 - Compensation

Instruction Manual CON 11/110 458.6 P5.0: Reset to factory defaults Program 5.0 reset functions lets you have an option to only reset all the me

Pagina 45 - TDS Conversion

Instruction Manual CON 11/110 11 INTRODUCTION Thank you for selecting the CON 11/110 Conductivity/TDS handheld meter. These meters are microproces

Pagina 46 - Figure 35: P2.0 –Viewing

Instruction Manual CON 11/110 468.6.2 P5.2: User Reset This reset function clears all the calibrations, memory and all other setup parameters tha

Pagina 47 - Figure 36: P3.0-Viewing

Instruction Manual CON 11/110 478.7 P6.0: Memory Clear Use this program to clear all memory values when you need to store a new series of values

Pagina 48 - 8.5 P4.0: Auto Off

Instruction Manual CON 11/110 488.8 P7.0: Temperature 8.8.1 P7.1: Setting the temperature coefficient The temperature coefficient is the amount o

Pagina 49 - Figure 38: P5.1-Clear all

Instruction Manual CON 11/110 498.8.2 P7.2: Setting the normalisation temperature Your meter will normalise its conductivity measurements to a st

Pagina 50 - Figure 39: P5.2-Clear

Instruction Manual CON 11/110 508.9 P8.0: Selection of cell constant Your meter lets you select a cell constant of K = 1.0, 10, or 0.1. Use a ce

Pagina 51 - 8.7 P6.0: Memory Clear

Instruction Manual CON 11/110 51 8.10 P9.0: Automatic and Single point Calibration You may select an automatic or manual calibration depending o

Pagina 52 - 8.8 P7.0: Temperature

Instruction Manual CON 11/110 528.10.2 P9.2: Selection of Single or Multi Point Calibration The single point calibration lets you have a single ca

Pagina 53 - Normalisation Temperature

Instruction Manual CON 11/110 539 CYBERCOMM PORTABLE DAS (FOR CON 110 ONLY) The DAS software is designed for pH 110 and CON 110 meters to allow yo

Pagina 54 - Selecting Cell Constant

Instruction Manual CON 11/110 54 Figure 47: Click on 'Browse' button and locate CD-ROM drive Figure 48 : Locate the CyberComm Portable

Pagina 55 - Automatic or Manual

Instruction Manual CON 11/110 55 Figure 49: Select ‘Portable Meters’ directory to go to CyberComm Portable. Figure 50 : Open ‘CyberComm Portabl

Pagina 56 - Multi point

Instruction Manual CON 11/110 22 DISPLAY AND KEYPAD FUNCTIONS 2.1 Display The LCD has a primary and secondary display. • The primary display sho

Pagina 57 - 9.1 System Requirements

Instruction Manual CON 11/110 56 Figure 51: Select "SETUP.EXE" program and click the OPEN button Figure 52: InstallShield Wizard dialo

Pagina 58 - DAS" directory

Instruction Manual CON 11/110 57 Figure 54: Key in your name and company name and click NEXT button Figure 55: To select another Destination Dir

Pagina 59

Instruction Manual CON 11/110 58 Figure 56: Creating a new program folder. Click on NEXT button. Figure 57: Click on NEXT button.

Pagina 60

Instruction Manual CON 11/110 59 Figure 58: The CyberComm Portable DAS program is fully installed. Click on FINISH button to end installation. 9.3

Pagina 61

Instruction Manual CON 11/110 60 Figure 60: The opening screen will appear as above 9.3.1 Buttons & Check-Box 1. Enable Connection - Click

Pagina 62

Instruction Manual CON 11/110 61 Figure 61: Under File Menu setting, you can change various parameters. Under ABOUT menu, details of Eutech Instru

Pagina 63

Instruction Manual CON 11/110 62 Figure 62: Communication Settings for computer's Com port. It must match with COM port settings on CyberScan

Pagina 64

Instruction Manual CON 11/110 639.4 Capturing And Printing Data Into Computer Using Data Acquisition After matching the Communication Settings b

Pagina 65

Instruction Manual CON 11/110 649.5 Trouble-shooting Guide Problem: Unable to PRINT When press PRINT key on CON 110 meter, nothing is printed. PO

Pagina 66

Instruction Manual CON 11/110 6510 ELECTRODE CARE AND MAINTENANCE Keep the conductivity electrode clean. Rinse the electrode twice, and gently swi

Pagina 67

Instruction Manual CON 11/110 32.2 Keypad A large membrane keypad with tactile feedback makes the instruments easy to use (Figure 2). Each button,

Pagina 68 - POSSIBLE CAUSES SOLUTIONS

Instruction Manual CON 11/110 6611 ERROR MESSAGES The following table provides a guideline to enable diagnosis of possible problems indicated by t

Pagina 69 - CONSEN91W/35608-50)

Instruction Manual CON 11/110 6712 TROUBLE SHOOTING Problem Cause Solution No display when ‘ON’ key is pressed a) Batteries not in place b) Ba

Pagina 70 - 11 ERROR MESSAGES

Instruction Manual CON 11/110 6813 LIST OF ACCESSORIES 13.1 Replacement Meter and Meter accessories Description Eutech Instruments Ordering Code

Pagina 71 - 12 TROUBLE SHOOTING

Instruction Manual CON 11/110 6914 FACTORY DEFAULT SETTINGS 14.1 CON 11 Factory Default Setting PROGRAM FUNCTION PARAMETER OPTING KEYS PARAMETER

Pagina 72 - 13 LIST OF ACCESSORIES

Instruction Manual CON 11/110 7014.2 CON 110 Factory Default Setting PROGRAM FUNCTION PARAMETER OPTING KEYS PARAMETER OPTIONS DEFAULT SETTING P 1

Pagina 73

Instruction Manual CON 11/110 7115 SPECIFICATIONS SPECIFICATIONS CON 11 CON 110 Conductivity Range 0 to 19.99 µS/cm, 20.0 to 199.9µS/cm, 200

Pagina 74

Instruction Manual CON 11/110 72Auto HOLD function Not available Selectable Averaging/Stability (READY) Selectable Selectable Inputs 6-pin

Pagina 75 - 15 SPECIFICATIONS

Instruction Manual CON 11/110 7316 CALIBRATION TIPS You only need one calibration for measurement throughout the entire range of the meter. If a r

Pagina 76

Instruction Manual CON 11/110 7418 CALCULATING TEMPERATURE COEFFICIENTS To determine the temperature coefficient of your sample solution use this

Pagina 77 - 16 CALIBRATION TIPS

Instruction Manual CON 11/110 7519 WARRANTY This meter is supplied with a three-year warranty, six-month warranty for electrode against significan

Pagina 78

Instruction Manual CON 11/110 4 MODEMEASCALENTERHOLDMRMISETUPRANGEOFFONMI MRSETUP MODEMEASCALENTERHOLDOFFONRANGEKeypad for CON 11 meter

Pagina 79 - 20 RETURN OF ITEMS

For more information on Eutech Instruments/ Oakton Instruments’ products, contact your nearest distributor or visi

Pagina 80

Instruction Manual CON 11/110 53 PREPARATION 3.1 Inserting the Batteries This meter is packaged with 4 “AAA” alkaline batteries required for its o