Instruction Manual αlpha-pH2000D
- 16 -
5.3 ORP-mV Calibration
If controller is configured for ORP-mV measuring you can perform a one-point calibration.
MEAS
CAL
CAL
HOLD
CAL
HOLD
mV
mV
R.mV
CAL
250
255
255
255
250
1. Enter Calibration mode as described in section 5.1. The display should read “CAL ORP”.
Note: If the display reads “CAL PH”, see section 6.9 for procedures on how to switch
from pH to ORP-mV readings.
2. Place sensor in the ORP solution.
3. Press the ENT key to start calibration. The “CAL”, buffer and electrode indicator appears on
the display. The upper and lower display shows the current mV output of the electrode without
any offset adjustment.
4. Press the or key to adjust the upper display mV value to match your exact reading.
5. Press the ENT key (or CAL key) to confirm. The controller calculates the correction factor and
returns to the Measurement mode. The “R.mV” annunciator (Relative mV) will be shown on the
display indicating that the value displayed is the relative mV value of the default measurement.
5.4 ORP -% Calibration
If controller is configured for ORP-% measuring you can perform a calibration at two points
using a low concentration sample (20%) and a high concentrate sample (80%). To calibrate
the controller for ORP %:
CAL
CAL
MEAS
59.0
%
%
%
CAL
CAL
HOLD
CAL
HOLD
C.1
C.2
1. Enter Calibration mode as described in section 5.1. The display should read “CAL ORP”.
Note: If the display reads “CAL PH”, see section 6.9 for procedures on how to switch
from pH to ORP-mV readings.
2. Press the ENT key to start calibration. The display prompts you to place the sensor in the high
concentration sample (relative value of 80%).
3. Place sensor in high concentration sample, then press the ENT key (or CAL key). The
electrode and CAL indicators flash. The unit automatically adjusts its reading to match the
solution value.
4. The display prompts you to place the sensor in the low concentration sample (relative value of
20%). Remove the electrode from the first standard, wash it well and then immerse it in the 20%
standard.
5. When the display stabilizes press the ENT key (or CAL key). The electrode and “CAL”
indicator flashes and the controller adjust its display to the 20% relative value and exits to the
Measurement mode.