Instruction Manual αlpha-pH2000D
- 29 -
6.9 Configuration (CNFG) sub-function
In this sub-function you configure the transmitter to suit your requirements.
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
mV
%
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
1
2
2
3
SETUP
HOLD
4
4
4
5
SETUP
HOLD
SETUP
HOLD
ENT
ENT
6
7
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
MEAS
pH
O
C
ENT
ENT
ENT
6
7
7
SY
SY
SY
1. Select the “CNFG” sub-function, and then press the ENT key.
2. Enabling/disabling the display backlight: press the or key to switch display backlight
on (BKLT ON) or off (BKLT OFF). Press the ENT key to confirm your selection.
3. Setting the backlight intensity (appears only If backlight is on): press the or key to set
backlight intensity (setting range: minimum 1 to maximum 4). Press the ENT key to confirm
your setting.
4. Selecting the unit of measurement: press the or key to select measurement mode:
- “PH SY” = pH measurement with symmetrical input
- “ORP mV SY” = ORP-mV measurement with symmetrical input
- “ORP % SY” = ORP-% measurement with symmetrical input
- Press the ENT key to confirm your selection.
5. Setting the alarm delay time in seconds: press the or key to set the alarm delay time
(setting range: 0 to 2000 seconds). Press the ENT key to confirm your setting.
Note: With this parameter you set the delay before the alarm relay activates when
the set point has been exceeded.