8 TECHNICAL SPECIFICATIONS
8.1 General Specification
(a) pH
pH Measuring Range -2.00 to 16.00 pH
Resolution 0.01 pH
Accuracy ± 0.01 pH
(b) mV
mV Measuring Range -1000 to 1000 mV
Resolution 1 mV
Accuracy ± 1 mV
(b) mV percentage
Percentage mV Measuring Range 0 to 100% (for -100 to 100 mV)
Resolution 0.1%
Accuracy ± 0.2 % (or ± 1 mV)
(c) Temperature
Temperature measuring range -10.0 to +125.0
o
C or +14.0 to +257.0
o
F for pt100/pt1000
-10.0 to +110.0
o
C or +14.0 to +230.0
o
F for NTC300
Resolution 0.1
o
C or
0.1
o
F
Relative accuracy 0.5
o
C or
1.0
o
F
Sensor NTC 300, Pt100 or Pt1000 (Jumper Selectable)
– 2 wire or 3 wire
Compensation Automatic (± 10 ˚C/ ± 18 ˚F offset adjustment) or Manual
8.2 Set-point and control function – Relay 1 & Relay 2
Function Switchable
- Selection in Control (CNTR) function of
Setup
Off
Limit Control
P/PI Control (Pulse length / Pulse frequency)
(a) Off
The control function of the controller is off SP1 & SP2 function will not appear in setup when Off is
selected in CNTR function to off mode
(b) Limit Control & P/PI Control
Relay contact selection for active mode Energise (EN) or
De-energise (DEEN)
Set Point 1 (SP1) -2.00 to 16.00 pH or 0 to 100% or -1000 to 1000mV
Set Point 2 (SP2) -2.00 to 16.00 pH or 0 to 100% or -1000 to 1000mV
Active mode selection (for SP1 & SP2) LO (Relay is active when measured value < set point)
HI (Relay is active when measured value > set point)
Hysteresis (for SP1 & SP2) 0.1 to 1 pH or 1 to 10.0%/ 10 to 100 mV
On (pickup) Delay (for SP1 & SP2) 0 to 2000 seconds
Off (Dropout) Delay (for SP1 & SP2) 0 to 2000 seconds
(c) P/PI Control only
Proportional Range 10 to 500 %
Integral Action Time (IAT) 0.0 to 999.9 minutes