EUTECH INSTRUMENTS ALPHA CON 100 CONDUCTIVITY CONTROLLERTRANSMITTER Manual de instrucciones Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 34
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
15
8.2 Setup mode overview
P1.0: Set Point 1
P1.1: select relay 1 set point value
P1.2: select relay 1 as low or high set point
P1.3: set relay 1 hysteresis value
P2.0: Set Point 2
P2.1: select relay 2 set point value
P2.2: select relay 2 as low or high set point
P2.3: set relay 2 hysteresis value
P3.0: Range
P3.1: select measurement range (with the corresponding cell constant value).
Eight measurement ranges are available.
P4.0: Temperature Data
P4.1: select ATC or MTC
P4.2: select temperature coefficient value
P4.3: select normalization temperature
P5.0: Conductivity Calibration points
P5.1: view conductivity value at which calibration was performed
P6.0 Electrode Properties
P6.1: view cell constant value after calibration
P6.2: view temperature offset value after calibration (only if in ATC mode)
P7.0 Reset to Factory Defaults
P7.1: select YES/NO to confirm reset decision
Vista de pagina 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 33 34

Comentarios a estos manuales

Sin comentarios