EUTECH INSTRUMENTS ALPHA CON 100 CONDUCTIVITY CONTROLLERTRANSMITTER Manual de instrucciones Pagina 23

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 34
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 22
20
8.5 Measurement Range Selection – P3.0
Setup program P3.0 is for selecting the range of measurement.
P3.1: select measurement range and corresponding cell constant
P3.
0
RNGE
P3.
1
2000
K
10
.
Press and keys together (ESCAPE) at anytime, to leave the Setup mode and
return to Measurement mode.
P3.1: Select Measurement Range
Set controller to the specific range of measurement. Ensure the cell you have
connected to the controller has the same cell constant as that stated in the range.
1. Key in the password “022” as per procedure in Section 5.
2. Press key until screen displays P3.0 and RNGE. Press ENTER.
3. The screen will scroll P3.1 and the selected range.
4. Press or keys to select appropriate range and cell.
Available ranges are as follows (Default range is highlighted):
Conductivity Range Cell Constant 4 mA current 20 mA current
0.00 – 20.00 µS/cm 0.1 0.00 µS/cm 20.00 µS/cm
0.0 – 200.0 µS/cm 0.1 0.0 µS/cm 200.0 µS/cm
0.0 – 200.0 µS/cm 1 0.0 µS/cm 200.0 µS/cm
0 – 2000 µS/cm 1 0 µS/cm 2000 µS/cm
0.00 – 10.00 mS/cm 1 0.00 mS/cm 10.0 mS/cm
0.00 – 20.00 mS/cm 1 0.00 mS/cm 20.0 mS/cm
0.0 – 200.0 mS/cm 1 0.0 mS/cm 200.0 mS/cm
0.0 – 200.0 mS/cm 10 0.0 mS/cm 200.0 mS/cm
5. Press ENTER to confirm (The set points 1 & 2, Conductivity and
Temperature calibration are reset to default values).
6. Press and keys together (ESCAPE) to return to P3.0. Press the and
keys together again, to return to measurement mode.
Vista de pagina 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 33 34

Comentarios a estos manuales

Sin comentarios