8.7 Viewing Conductivity Calibration Data – P5.0
Program 5 is a “view only” option, which displays the value at which calibration was
performed.
P5.1: view Calibrated conductivity value (if no calibration performed for this range,
controller displays ‘- - - - ‘).
These parameters will change each time you recalibrate the controller.
P5. 0
C L
P5. 1
CAL
300
Press and keys together (ESCAPE) at anytime, to leave the Setup mode and
return to Measurement mode.
P5.0: Viewing Calibrated Conductivity data
1. Key in the password “022” as per procedure in Section 5.
2. Press key until screen displays P5.0 and CAL. Press ENTER.
3. Screen will scroll P5.1, ‘CAL’, then Calibrated value (in uS/cm or mS /cm).
4. Press and keys together (ESCAPE) twice, to return to measurement
mode.