EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY CONTROLLERTRANSMITTER (REVISION 3) Manual de instrucciones Pagina 14

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 40
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 13
14
6 TDS CALIBRATION
IMPORTANT: When Calibration mode is entered, controller automatically goes into a
“HOLD” mode where the 4-20 mA output freezes and relays are de-activated (if it was in
an activated condition). Upon return to measurement mode, both 4-20mA output and
relay activities resume, depending on settings.
The αlpha TDS 200 controller includes 8 total dissolved solids measurement ranges.
One-point slope calibration is possible, in each range.
Choose fresh standard solutions, whose value is reasonably close to the measurement
value.
Before calibration, make sure electrode is clean. Use iso-propyl alcohol followed by
thorough rinsing in distilled water to clean electrode.
Important: To achieve a successful TDS calibration, two conditions must be satisfied:
1. Difference between measured value of standard solution, and actual value of
standard solution, must be within ± 40 %.
(e.g. if value of standard solution is 500 ppm, then allowable measured
values by controller must be within 700 ppm {+ 40%} and 300 ppm{- 40%});
and
2. the lowest possible value must be at least 10% of full scale
(e.g. if range of 0 to 1000 ppm is selected, min. measured value at
calibration, must be at least 100.0 ppm).
Vista de pagina 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 39 40

Comentarios a estos manuales

Sin comentarios