Press and keys together (ESCAPE) at anytime, to leave the Setup mode and return
to Measurement mode.
P8.0: Controller reset
1. Key in the password “022” as per procedure in Section 5.
2. Press key until screen displays P8.0 and RST. Press ENTER.
3. Screen will scroll P8.1, then ‘No’.
4. Press or keys to toggle between YES and NO. Press ENTER to confirm
selection.
5. Press and keys together (ESCAPE) to return to P8.0. Press and keys
together again, to return to measurement mode.