8.8 Viewing TDS / Temperature Electrode Data – P6.0
Program 6 has two “view only” options that let you check the electrode parameters for
diagnostic purposes.
P6.1: view revised cell correcting factor value of electrode
P6.2: view temperature probe offset (ATC on only)
These parameters will change each time you recalibrate the controller.
P6.0
elec
P6.1
P6.2
t.ofs
0.0
1000
.
FCTR
Press and keys together (ESCAPE) at anytime, to leave the Setup mode and return
to Measurement mode.
P6.1: Viewing TDS electrode data
1. Key in the password “022” as per procedure in Section 5.
2. Press key until screen displays P6.0 and ELEC. Press ENTER.
3. Screen will scroll P6.1, ‘FCTR’ then cell correction factor. Each time a calibration is
performed, the value of the cell correction factor will be updated.
4. Press and keys together (ESCAPE) to return to P6.0 or press ENTER to view
Temperature electrode data.
P6.2: Viewing Temperature electrode data
5. Follow procedure from above (P6.1) until step 3 and press ENTER.
6. Screen will scroll P6.2, ‘T.OFS’ (Temperature Offset), then amount of offset (in
o
C).
7. Press ENTER or press and keys together (ESCAPE) twice, to return to
measurement mode.