4. If necessary, press the MI/ key to select the next
“memory location” screen; press the MR/ key to
select the previous “memory location” screen.
5. Repeat steps 2 to 5 to review additional stored data
sets.
6. To exit Memory Recall, press the MEAS key to
return to the Measurement mode.
NOTES:
Readings stored in memory are retained even if the units are
turned off. To erase all readings stored in memory, use the
SETUP mode P1.0 on page 25.