EUTECH INSTRUMENTS PH 6 PLUS PH MV TEMP Manual de instrucciones Pagina 33

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 64
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 32
Instruction Manual CyberScan pH 1100/ 2100
33
10.2.6 pH Calibration Due Alarm Setup program P1.4
Use this program to set a reminder when the next calibration is due.
Choose from 1 to 365 days from last calibration.
1. Enter pH/Temperature setup mode as described in steps 1-4 in
10.2.1 Entering pH/temperature setup mode (page 30).
2. Use /MI to scroll to program pHCalDue Alarm: P1.4. See
Figure 91.
3. Press ENTER to enter program pHCalDue Alarm: P1.4.
4. Press >/CAL to toggle between Cal Alarm Due:On and Off. The
selection will flash. See Figure 92.
5. Press ENTER to confirm selection.
If you selected On, continue to step 6
If you select Off, pressing ENTER returns to the setup mode.
Press /MI to continue to the next setup program, or press /MR
to go to the previous program. Press EXIT to return to the pH
measurement mode.
6. If you selected On in step 5, the display reads Key in Cal Due
Period (1-365 days) =. Key number of days using the numeric
keypad. If you make a mistake, use the /MI or /MR keys to
highlight mistake, then re-enter. See Figure 93.
7.
Press ENTER to confirm number of days to next calibration,
screen flashes Sys Updating, Please wait... and returns to the
setup mode. Press /MI key to continue to the next setup
program, or press /MR to go to the previous program. Press
EXIT to return to the pH measurement mode.
Keyin Cal Due Period
(1-365 days) = 1
Figure 91
Figure 92
Figure 93
Cal Alarm Due: On
Select & Press ENTER
Off
pHCalDue Alarm : P1.4
Press ENT. to Select
Vista de pagina 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 63 64

Comentarios a estos manuales

Sin comentarios