Instruction Manual CyberScan pH 1100/ 2100
36
10.2.9 pH calibration data reset setup program P1.7
Use this program to erase all pH calibration data.
1. Enter pH/Temperature setup mode as described in steps 1-4 in
Section 10.2.1 Entering pH/temperature setup mode (page 30).
2. Press /MI to scroll to program pH Cal Reset: P1.7. See
Figure 105.
3. Press ENTER to enter program pH Cal Reset: P1.7.
4. Press >/CAL to toggle between pH Reset: Yes and No. The
selection will flash. See Figure 106.
5. Press ENTER to confirm selection.
If you selected Yes, the screen flashes pH being Reset,
Please wait... See Figure 107. The meter then returns to
setup mode.
If you select No, the screen flashes Reset Aborted and
then returns to the setup mode. See Figure 108.
This is the last pH setup program. Press /MR to go to the previous
program. Press /MI to go back to the first pH setup program. Press
EXIT to return to the pH measurement mode.
Reset Aborted
pH being Reset
Please wait. . .
pH Reset: Yes No
Select & Press ENTER
pH Cal Reset :P1.7
Press ENT. to Select
Figure 105
Figure 106
Figure 107
Figure 108