10.2.9 pH calibration data reset setup program P1.7
Use this program to erase all pH calibration data.
1. Enter pH/Temperature setup mode as described in steps 1-4 in
Section 10.2.1 Entering pH/temperature setup mode (page 30).
2. Press ∆/MI to scroll to program pH Cal Reset: P1.7. See
Figure 105.
3. Press ENTER to enter program pH Cal Reset: P1.7.
4. Press >/CAL to toggle between pH Reset: Yes and No. The
selection will flash. See Figure 106.
5. Press ENTER to confirm selection.
• If you selected Yes, the screen flashes pH being Reset,
Please wait... See Figure 107. The meter then returns to
setup mode.
• If you select No, the screen flashes Reset Aborted and
then returns to the setup mode. See Figure 108.
This is the last pH setup program. Press ∇/MR to go to the previous
program. Press ∆/MI to go back to the first pH setup program. Press
EXIT to return to the pH measurement mode.