10.4.6 Ion mode setup program P3.4
Use this program to select high or low ion resolution mode. High
resolution has ±500 mV value, low has ±1850 mV value.
1. Enter ion setup mode as described in steps 1-4 in Section
10.4.1 Entering ion setup mode (page 43).
2. Use ∆/MI key to scroll to program Ion Mode Setup: P3.4. See
Figure 144.
3. Press ENTER to enter program Ion Mode Setup: P3.4.
4. Press >/CAL to toggle between High or Low. The unit
selected will flash. See Figure 145.
5. Press ENTER to confirm selection and return to the setup
mode.
6. Press ∆/MI key to continue to the next setup program, or
press ∇/MR to go to the previous program. Press EXIT to
return to the ion measurement mode.
Resolution: High Low
Select & Press ENTER
Ion Mode Setup : P3.4
Press ENT. to Select
Figure 144
Figure 145