Instruction Manual PC 510
7
5.3 pH calibration
NOTE: We recommend that you perform at least a 2-point calibration using standard buffers that
bracket (one above and one below) the expected sample range.
5.3.1 Preparing for pH calibration
This meter is capable of up to 5-point pH calibration to ensure accuracy across the entire pH range of
the meter. Select from the following buffer options:
USA: pH 1.68, 4.01, 7.00, 10.01 and 12.45
NIST: pH 1.68, 4.01, 6.86, 9.18 and 12.45
The meter automatically recognises and calibrates to these standard buffer values, which makes pH
calibration faster and easier.
NOTE:
Selection of USA or NIST buffer standards must be done prior to calibration. Refer to Section
8.4 on P3.3 on page 29.
5.3.2 Before starting
Be sure to remove the protective electrode storage bottle or rubber cap
of the probe before calibration or measurement. If the electrode has
been stored dry, hydrate the probe in tap water for 10 minutes before
calibrating or taking readings to saturate the pH electrode surface and
minimise drift.
Wash your probe in de-ionised water after use, and store in electrode
storage solution. If storage solution is not available, use pH 4.01 or
7.00 buffer for short term storage. DO NOT store electrode in distilled
or de-ionised water.
5.3.3 To calibrate pH
1. If necessary, press the MODE key to select pH mode. The pH
indicator appears in the upper right hand corner of the display.
2. Rinse the probe thoroughly with de-ionised water or rinse solution.
Do not wipe the probe as this causes a build-up of electrostatic
charge on the glass surface.
3. Dip the probe into the standard calibration buffer. The end of the
probe must be completely immersed into the sample. Stir the probe
gently to create a homogeneous sample.
NOTE
: The temperature element is in the conductivity cell. For
temperature compensated readings, dip the conductivity cell or ATC
probe into the calibration buffer as well.
4. Press CAL/MEAS to enter pH calibration mode. The CAL indicator
will be shown. The primary display will show the measured reading
while the smaller secondary display will indicate the pH standard
buffer solution.
5. Wait for the measured pH value to stabilise. If the READY indicator
has been activated through the Setup, the READY appears when
the reading is stable.
6. Press ENTER to confirm calibration. The meter is now calibrated to
the current buffer. The lower display automatically scrolls through
the remaining buffer options.
If you are performing multi-point calibration, go to step 7.
MEAS
pH
ATC
°C
7.16
22.3
CAL
pH
pH
7.16
7.00
C
A
L
M
E
A
S
CAL
pH
pH
READY
7.16
7.00
CAL
pH
pH
READY
7.00
7.00
E
N
T
E
R
R
A
N
G
E
Figure 6: pH calibration