Instruction Manual PC 510
3
S
E
T
M
O
D
E
H
O
L
D
O
N
/
O
F
F
E
N
T
E
R
R
A
N
G
E
C
A
L
M
E
A
S
Figure 4: Keypad
ON/OFF
Powers the meter on or off. When meter is switched on, it starts in the mode the meter
was last in when powered off.
SET
Enters advanced setup mode. SETUP mode lets you customise meter preferences and
defaults, and view calibration and probe data.
MODE
Measurement: Press MODE to toggle between pH, mV, conductivity and TDS.
Calibration: In calibration mode, press MODE to access temperature calibration.
HOLD
Freezes the measured reading. To activate, press HOLD while in measurement mode. To
release, press HOLD again.
Note: When auto endpoint feature is switched on, meter automatically holds reading after
5 seconds of stability. The HOLD indicator appears on the display. Press HOLD to release
auto endpoint feature.
CAL/MEAS
Toggles between Calibration and Measurement mode. Example: If you are in pH
measurement mode, press CAL/MEAS to enter pH calibration mode.
Note: Temperature calibration is available from pH, TDS or Conductivity calibration mode.
In Setup mode: Press CAL/MEAS to return to main menu from sub menus. Press
CAL/MEAS again to return to measurement mode from main menu.
ENTER/RANGE
ENTER: Press to confirm values in Calibration mode and to confirm selections in Setup
mode.
RANGE: Press to switch to manual ranging in Conductivity or TDS mode.
Press in Setup mode to scroll up through su/jointfilesconvert/437882/bgroups. Also lets you increase the values
in the conductivity, TDS and temperature calibration modes.
Press in Setup mode to scroll down through su/jointfilesconvert/437882/bgroups. Also lets you decrease the
values in the conductivity, TDS and temperature calibration modes.