EUTECH INSTRUMENTS CYBERSCAN CON 1500 CONDUCTIVITY METER Manual de instrucciones Pagina 12

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 11
Instruction Manual PC 510
8
If you are performing one-point calibration, go to step 9.
7. Rinse the electrode with de-ionised water or rinse solution, and
place it in the next pH buffer.
8. Follow steps 5 to 7 for additional calibration points.
9. When calibration is complete, press CAL/MEAS to return to pH
measurement mode.
NOTE
: To exit from pH calibration mode without confirming calibration,
DO NOT press ENTER in step 6. Press CAL/MEAS instead.
If the selected buffer value is not within ±1.0 pH from the measured pH
value: the electrode and buffer icon blink and the ERR annunciator
appears in the lower left corner of the display.
To limit the number of pH buffer values available during calibration,
see section 8.4 Setup P3.2 on page 29.
Figure 7: Next point calibration
for pH 4.01
CAL
pH
pH
READY
4.15
4.01
CAL
pH
pH
READY
4.01
4.01
E
N
T
E
R
R
A
N
G
E
Figure 8: Err message and
electrode icon will appear if
in
co
rr
ect
bu
ff
e
r i
s
used
CAL
pH
pH
ERR
0.64
1.68
Vista de pagina 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 55 56

Comentarios a estos manuales

Sin comentarios