Instruction Manual PC 510
17
6.2.3 Setting a manual temperature compensation value
To use manual temperature compensation, you need to enter the temperature value of your process
into the meter. This is the value at which reading will manually temperature compensate. You can
select any temperature between 0 and 100 °C (32 to 212 °F). Default value is 25 °C (77 °F).
To select a manual temperature compensation value
1. Switch the meter on. Press the MODE key to select conductivity or
TDS measurement mode.
2. If necessary, select ATC off as described in P7.3 on page 35. The
ATC indicator will not appear on the display.
3. Press the CAL/MEAS key to enter conductivity or TDS calibration
mode. The CAL indicator will appear above the primary display.
4. While in conductivity (or TDS) calibration mode, press the MODE
key to enter temperature calibration mode. The primary and
secondary displays show the last set temperature value.
5. Check the temperature of your sample using an accurate
thermometer.
6. Press the or key to offset the temperature to the measured
value from step 5.
7. Press ENTER to confirm the selected temperature and to return to
the conductivity measurement mode. See Figure 18.
The meter will now compensate conductivity or TDS readings for
manually set temperature.
NOTES
: To exit this program without confirming the manual
temperature compensation value, DO NOT press ENTER in step 7.
Press CAL/MEAS instead.
Setting the manual temperature compensation value for conductivity
will change the manual temperature compensation value for pH to the
same value, and vice versa.
6.2.4 Taking Measurements (Conductivity or TDS)
To take readings:
1. Rinse the probe with de-ionised or distilled water before use to
remove any impurities adhering to the probe body. Shake or air
dry. To avoid contamination or dilution of your sample, rinse probe
with a small volume of your sample liquid.
2. Press ON to switch meter on.
3. Press the MODE key to select conductivity or TDS measurement
mode. The MEAS annunciator appears on the top center of the
LCD. The ATC indicator appears in the lower right hand corner to
indicate Automatic Temperature Compensation.
NOTE
: For conductivity manual temperature compensation, you must de-activate the temperature
sensor built into the conductivity probe and set a manual temperature compensation. See page 14 for
more information. The ATC indicator will disappear from the display.
Figure 19: Manual Temperature
Compensation
Figure 20: During measurement
CAL
°C
25.0
25.0
MEAS
°C
256
30.0
READY
E
N
T
E
R
R
A
N
G
E
°C
30.0
25.0
CAL
MEAS
READY
ATC
°C
345
21.3
µ
S