Transmitter Cond/Res 2-Channel M300 25
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
7.2.1 One-point Sensor Calibration
(Display reflects typical Conductivity Sensor calibration)
Select 1 point Calibration by pressing the or key followed by the [ENTER] key.
Enter the Value of calibration Point 1 and then press the [ENTER] key to start calibration.
The value in the 2
nd
text line is the actual measured value from the sensor prior to calibration.
After the calibration the Multiplier or slope calibration factor ‘M’ and the Adder or offset calibration
factor ‘A’ are displayed.
Select Yes to save the calibration values and the successful Calibration is confirmed on the
display.
7.2.2 Two-point Sensor Calibration
Enter Sensor Calibration mode as described in section 7.1.
Select 2 point Calibration by pressing the or key followed by the [ENTER] key.
CAUTION: Rinse sensors with a high-purity water solution between calibration points to prevent
contamination of the reference solutions.
Enter the Value of Point 1 and press the [ENTER] key. Place the sensor into the second reference
solution.
Enter the Value of Point 2 and press the [ENTER] key to start the calibration.
After the calibration the Multiplier or slope calibration factor ‘M’ and the Adder or offset calibration
factor ‘A’ are displayed.
Select Yes to save the net calibration values and the successful Calibration is confirmed on the
display.
7.3 Sensor Temperature Calibration
Enter Sensor Calibration mode as described in section 7.1 and select Temperature.