Transmitter Cond/Res 2-Channel M300 37
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
8.5.2 Clean
Configure the Relay to be used for the cleaning cycle.
The Cleaning Interval can be set from 0.000 to 999.9 hours. Setting it to 0 turns the clean cycle
off. The cleaning time can be 0 to 9999 seconds and must be smaller than the Cleaning
Interval.
Select the desired Relay state: Normal or Inverted.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will dis-
card the entered values and return to the measurement display screen, selecting Yes will save
changes made.
8.6 Display
(PATH: Menu/Configure/Display)
Enter configuration mode as described in Section 8.2.
This Menu allows for the configuration of the values to be displayed and also the configuration
of the Display itself.
8.6.1 Measurement
The Display has 4 lines. Line 1 on top and Line 4 on the bottom.
Select the values (Measurement a, b, c or d) to be displayed on each line of the display.
Select the ‘Error Display’ mode. If this is set to ‘On’ when an alarm has occurred, the message
“Failure – Press Enter” will be displayed on Line 4 when an alarm occurs in the normal
Measurement mode.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard
the entered values, selecting Yes will make the entered values the current ones.