Enter the desired value(s) for the Setpoint and press [ENTER]
This screen provides the option to configure a setpoint to be active on an over range condition.
Select the setpoint (1 thru 4) and ‘Yes’ or ‘No’. Select the desired relay that will activate when the
setpoint alarm condition is reached.
Over Range
Configure whether an over range should also be alarmed and which relay should be used. Once
configured, the selected relay will be activated if a sensor over-range condition is detected on the
assigned input channel.
Delay
Enter the delay time in seconds. A time delay requires the setpoint to be exceeded continuously
for the specified length of time before activating the relay. If the condition disappears before the
delay period is over, the relay will not be activated.
Hysteresis
Enter the hysteresis as a percentage-value. A hysteresis value requires the measurement to
return within the setpoint value by a specified percentage before the relay is deactivated.
For a high setpoint, the measurement must decrease more than the indicated percentage below
the setpoint value before the relay is deactivated. With a low setpoint, the measurement must
rise at least this percentage above the set-point value before the relay is deactivated. For exam-
ple, with a high setpoint of 100, when this value is exceeded, the measurement must fall below
90 before the relay is deactivated.
State
Relay contacts are in normal state until the associated setpoint is exceeded, then the relay is
activated and the contact states change.
Select ‘Inverted’ to reverse the normal operating state of the relay (i.e. Normally open contacts
are in a closed state, and normally closed contacts are in an open state, until the setpoint is
exceeded). ‘Inverted’ relay operation is functional when power is applied to the M300 trans-
mitter.
Pressing the [ENTER] key again will bring up the Save Changes dialog Selecting No will dis-
card the entered values and return to the measurement display screen, selecting Yes will save
changes made.