Transmitter Cond/Res 2-Channel M300 39
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
8.7 Hold Analog Outputs
(PATH: Menu/Configure/Hold Outputs)
Enter configuration mode as described in Section 8.2.
The Digital Input used to remotely control the Hold function is configured with this Menu.
Initiating a hold condition will maintain the analog signal output and relay status at the value/
state at the time the Hold is initiated, for as long as the Hold state is maintained. In addition,
the USB output will be held if the USB Hold option is set to “Last Values”. The USB Hold feature
is set to “Off” by default. Refer to section 9.2 for more information on the USB settings.
Analog output and relay status will not be held if ‘No’ is selected. If ‘Yes’ is selected, Outputs will
be held depending on the status of the selected Digital Input. Digital Input choices are ‘High’,
‘Low’ or ‘Off’. All analog outputs and relay status will be held if the Digital Input is in the selected
state. If ‘Off’ is selected as Digital Input status, the Digital Input is inactive and the Hold status
will not be triggered through an external signal, although the outputs will be held during confi-
guration or calibration procedures as long as the Hold Outputs option is ‘Yes’.