Transmitter Cond/Res 2-Channel M300 40
© 05/07 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Cond/Res 2-Channel M300
Printed in Switzerland 52 121 318
9 System
(PATH: Menu/System)
While in Measurement mode press the key. Press the or key to navigate to ‘System’ –
Menu and press [ENTER]. Enter the security code ‘xxxxx’ if necessary (see section 9.3) and
press [ENTER] to confirm the code.
Refer to section 3.3.2 for information on use of the navigation keys.
9.1 Set Language
(PATH: Menu/System/Set Language)
This Menu allows the configuration of the Display language.
The following selections are possible: English, French, German, Italian and Spanish.
Pressing the [ENTER] key will bring up the Save Changes dialog.
9.2 USB
(PATH: Menu/System/USB)PATH: Menu/System/USB)Menu/System/USB)
This menu allows configuration of the USB hold function.
This may be set to either Off or Last Values. An external host device may poll the M300 for data.
If the USB Hold is set to Off, current values are returned. If the USB Hold is set to Last Values, the
values present at the time the hold condition was established are returned.
Details of USB functions and communication protocols are covered in separate documentation.
Press [ENTER] to bring up the Save Changes dialog.