Instruction Manual PCD 650
42
Ensure that you use new conductivity standard solutions or sachets during
calibration. Do not reuse standard solutions as it may be contaminated and
affect the calibration and accuracy of measurements. Use fresh calibration
solution each time you calibrate your meter. Keep solutions in a dry and cool
environment if possible.
For DO(%) and DO (mg/L) Calibration
Before starting calibration, make sure you are in the correct measurement
mode and in the correct calibration sequence. The temperature and the %
Saturation calibration must be done first before attempting to do the
mg/L (ppm) Concentration calibration.
Rinse the probe well in the de-ionized (DI) water or rinse solution and wipe
the probe carefully taking care of the membrane.
Calibrate the meter in all the modes to ensure the highest accuracy
throughout the DO measurement range. In % Saturation, the meter is able to
perform either a one point calibration or a 2 point calibration. For one point
calibration, it is recommended that you perform a 100% Saturation calibration
in saturated air. If you opt for 2 point calibration, you can calibrate for 100%
Saturation in saturated air and 0% Saturation using a zero oxygen solution.
All new calibration values will automatically override the existing data. It is
recommended to calibrate the meter periodically and or if it is suspected to
be inaccurate.
Always rinse the probe with either DI water or rinse solution before and after
each calibration/sample measurement. When calibrating in air, make sure
that any water droplets from the probe’s membrane are removed.
3.1.7 Accessing Calibration mode
From measurement mode, press CAL (F2) key. The meter goes to
corresponding calibration mode, based on the selected measurement mode.
If the meter is password protected, you will be prompted to enter password.
(Refer the below section)
3.1.8 Accessing Calibration mode when password protection enabled
Follow the steps below to access the calibration mode, when password
protection is enabled (Refer page 80).
1. Make sure you are in measurement mode. If required, press MODE (F3) to
switch to the measurement mode for which you wish to perform calibration.
2. Press CAL (F2) to go to calibration mode. Login Password screen appears
(Figure 39). The meter expects the 5-digit password specified in system
setup. (Refer ’System Settings – Page 5’ in page 80)