Instruction Manual PCD 650
68
To View calibration Report:
1. From % saturation measurement mode, press left or right arrow key to
navigate to other available functions until you see REPO function in the
LCD.
2. Press REPO (F2) key. The calibration report is shown in the display.
3. Press PRIN (F2) to transfer the calibration report to computer through IrDA.
(Refer page 37 to print data).
Figure 61: DO Report in % saturation mode
Function Keys available in saturation report screen:
PRIN (F2)
Transfers calibration report to Computer through IrDA
NEXT (F3)
Goes to saturation measurement mode
ESC (F4)
Goes to saturation measurement mode
ENTER
(Not functional)
(Not functional)