Instruction Manual PCD 650
56
Figure 49 : Conductivity calibration Screen
7. The upper display shows the conductivity reading of the solution with
respect to previous calibration (if any) & lower display shows the
conductivity reading of the solution without any calibration. Use &
keys to adjust the upper display of the conductivity value of your custom
solution.
8. Press ENTER key to confirm the calibration.
9. Press ESC (F4) to go to measurement mode.
10. As for the case of multi point calibration, rinse the probe with de-ionized
water, and repeat step 1 to 7 for every calibration using the desired
calibration solutions until all points have been calibrated.
Function Keys available in Conductivity Calibration mode:
TEMP (F1)
Goes to temperature calibration
CLR-C (F2)
Clears previous calibration (if any) after ENTER key is pressed in
confirmation screen
NEXT (F3)
Shows calibration report
ESC (F4)
Exits from calibration and goes back to conductivity measurement
mode
ENTER
Confirms the calibration
Increases/decreases conductivity cell reading
(Not functional)
3.6.2 Automatic calibration (For Conductivity Calibration)
In the Automatic Calibration mode, the meter is capable of accepting either
single-point or up to 5 points for multi-point calibration with maximum of 1 point
per specific measurement range. Select automatic calibration mode in the
Conductivity Setup screen. See ‘Conductivity Settings – Page 1’ on page 86
for the set up procedure. Press (F2) to start calibration and the meter displays
cell adjustment screen as mentioned in the above section.
Note: If you wish to
completely re-calibrate
the meter, you need to
clear previous
calibration data. Press
CLR-C (F2) key to
clear previous
calibration. The meter
shows you
confirmation screen.
Press ENTER key to
confirm deleting
p
revious calibration.