Instruction Manual PCD 650
60
8. Rinse the probe with de-ionized water. Place it in the next solution and
follow the steps 2 to 7 to calibrate other points, if ‘MULTI’ point calibration
is selected.
9. Press ESC (F4) to go to measurement mode. The meter shows the
resistivity reading of the solution in measurement mode.
3.7.1 Resistivity Calibration Report
Calibration report gives you information on calibration. It includes date &
time on which the last calibration was done and the offset.
To View resistivity Report:
1. From resistivity measurement mode, press left
or right arrow key to
navigate to other available functions until you see REPO function in the
LCD
2. Press REPO (F2) key. The calibration report is shown in the display [Figure
53].
3. Press PRIN (F2) to transfer the calibration report to the computer.( Refer
page 37 to print data)
Figure 53 : Resistivity report screen
Note: The function of the keys available on the resistivity cell
constant/calibration/ report screen is same as previously described in
conductivity calibration.