
Instruction Manual PCD 650
61
3.8 Salinity Calibration
Use a standard solution of known value for calibration. Use the single-point
standardization option to measure samples that close in value and multi-point
standardization to measure wide range of samples.
1. Switch on the meter. Make sure the meter is in salinity measurement mode.
2. Press CAL (F2) to go to calibration mode.
3. The meter shows ‘Calibration-Rinse Electrode’ screen [Figure 48-(a)] few
seconds and then shows the cell constant adjustment screen [Figure 54-(a)].
(a) (b)
Figure 54 : Salinity Cell constant & salinity calibration
4. The upper display of the cell constant adjustment screen [Figure 54-(a)] shows
the salinity cell constant of your solution & lower display shows the measured
value of the solution with adjusted cell constant. Press
& keys to adjust
the cell constant for your custom solution in upper display.
5. Press ENTER key to confirm the entered value and to go to salinity calibration
screen.
6. The screen shows two readings [Figure 54-(b)]. The upper display shows the
salinity reading of the solution with respect to previous calibration (if any) &
lower display shows the salinity reading of the solution without any calibration.
Use & keys to adjust the upper display of the salinity value of your
custom solution.
7. Press ENTER key to confirm the entered value.
Note: If the meter is password protected, you will be prompted to enter a
password. Refer ‘Accessing Calibration mode when password protection
enabled’ in pa
e 42.
Comentarios a estos manuales