Instruction Manual PCD 650
79
4.2.4 System Settings – Page 4
Figure 71: System Settings - Page 4
This page allows you to set wireless serial data communication related parameters.
Parameter Description Factory Default
Print Mode
IrDA – Sets serial data communication protocol to
IrDA
LED – Sets serial data communication protocol to
RS232C
MEM- Logs data to meter’s memory.
IrDA
Data Format
CyberComm – Select this format if you use
CyberComm Data Acquisition Software (DAS)
TEXT – Select this format if you use any other method
(such as Windows
®
Hyperterminal)
This parameter is used when downloading data from
the meter through IrDA
CyberComm
Current Data
Set
TIMED – Prints measurement data continuously at the
interval specified in ‘INTERVAL’ parameter below.
SINGLE – Prints only the currently measured reading
This parameter applies when PRIN key is pressed
from measurement mode to send the currently
measured readings to the computer.
TIMED
Interval
(3 Sec Step)
Time interval at which the meter should send currently
measured data to the printer/CyberComm/PCD
Acceptable range : 3 sec to 600 sec (in 3 sec steps)
(This parameter is applicable when ‘CURRENT DATA
SET’ is set to ‘TIMED’ and this is not editable when
‘CURRENT DATA SET’ is set to ‘SINGLE’)
9 Sec
Fixed Setting
Indicates serial communication settings in the format
of ‘Baud rate, Data bits-Parity bits-Stop bits’. This
parameter is not editable.
2400 8-N-1