Instruction Manual PCD 650
49
Figure 43: pH Calibration Report
Example: In the given pH calibration report (Refer Figure 43), 4-point calibration has
been done for USA buffer group (pH 1.68, pH 4.01, pH 7.00 & pH 12.45).
The slope is 93 for the range enveloped from pH 1.68 to pH 4.01.
The slope is 88 for the range enveloped from pH 4.01 to pH 7.00.
The slope is 94 for the range enveloped from pH 7.00 to pH 12.45 where pH
10.01 calibration point has been skipped.
Function Keys available in pH calibration report screen:
PAGE (F1)
Goes to the second page of the calibration report
BACK (F1)
Goes to the first page of the calibration report
PRIN (F2)
Transfers calibration report to Computer through IrDA
NEXT (F3)
Goes to pH measurement mode
ESC (F4)
Goes to pH measurement mode
ENTER
(Not functional)
(Not functional)
3.3.4 Average Slope Indicator of pH Probe
Each time you perform pH calibration; the meter calculates the average slope of your
probe and graphically indicates it in the header section of the LCD screen.
Figure 44 : Average slope indicator