Instruction Manual PCD 650
58
Note: If the user has selected ‘SINGLE’ point calibration and ‘AUTO’
calibration mode, the meter will go to measurement mode after step 4.
3.6.3 Calibration Report
Calibration report gives you detailed information on calibration. It includes date
& time, calibration points, range, factor and cell constant.
To View calibration Report:
1. From conductivity measurement mode, press left or right arrow key to
navigate to other available functions until you see REPO function in the
LCD.
2. Press REPO (F2) key. The calibration report is shown in the display.
3. Press PRIN (F2) to transfer the calibration report to computer through IrDA.
(Refer page 37 to print data).
Figure 51: Conductivity Calibration Report
Function Keys available in conductivity calibration report screen:
PRIN (F2)
Transfers calibration report to Computer through IrDA
NEXT (F3)
Goes to conductivity measurement mode
ESC (F4)
Goes to conductivity measurement mode
ENTER
(Not functional)
(Not functional)
Note: Auto calibration can also be done for electrode with cell constant 0.4 – 2.